Sensors & Analyzers

O₂ Purity Analyzers

Continuous Online Monitoring — Zirconia or Electrochemical, 0–25% O₂ Range

Key Features

Why This Accessory Matters

01

Real-Time Purity Display

Continuous O₂ concentration readout with automatic alarm at user-set threshold.

02

Dual Sensor Options

Zirconia (fast response, long life) or Electrochemical (low cost, no reference gas).

03

Automatic Alarm

Adjustable high/low O₂ alarm with relay output for automatic product diversion.

04

Remote Output

4–20 mA isolated output and Modbus RTU for integration with PLC/DCS.

Technical Specifications

Detailed Specifications

Measurement Range0–1%, 0–5%, 0–25% O₂ (selectable)
Sensor TypesZirconia (ZrO₂) or Electrochemical
Accuracy±1% of full scale (zirconia), ±2% (electrochemical)
Response Time< 15 seconds (T90)
Display4-digit LED or 3.5" TFT with bar graph
Output4–20 mA (isolated), Modbus RTU, 2x relay
CalibrationAuto-cal with ambient air (20.9% O₂)
Sample Flow0.1–1 L/min
FAQ

Frequently Asked Questions

Which sensor type should I choose for my PSA nitrogen generator?

For most industrial applications, electrochemical sensors offer the best value. Choose zirconia if you need faster response time, longer sensor life (3–5 years vs 1–2 years), or are measuring very low O₂ concentrations.

How often should I calibrate the analyzer?

Monthly calibration is recommended for electrochemical sensors. Zirconia sensors require calibration every 3 months. Auto-calibration with ambient air simplifies the process.